- No Barcodes: A Basic UDI-DI appears in EUDAMED, Declarations of Conformity, Notified Body certificates, and technical files—never on a packaging barcode. (To decode scannable UDI barcodes, use UDIParser.com.)
- 25-Character Ceiling: Under EU MDR Annex VI, the complete Basic UDI-DI including check characters must not exceed 25 characters.
- Bulk EUDAMED Management: Portfolios with 25+ Basic UDI-DI families benefit from dedicated EUDAMED XML & RIM platforms to avoid manual portal errors.
Basic UDI-DI vs. UDI-DI vs. UDI-PI
| Attribute | Basic UDI-DI | UDI-DI | UDI-PI |
|---|---|---|---|
| Role | Device family key in EUDAMED | Packaging/SKU trade item key | Batch, lot, serial, or expiration |
| Printed in Barcode? | No (strictly prohibited) | Yes | Yes |
| GS1 Terminology | Global Model Number (GMN) | GTIN-14 (AI 01) | AI 17 / 11 / 10 / 21 |
Frequently Asked Questions
What is a Basic UDI-DI under EU MDR and IVDR?
A Basic UDI-DI is the primary access key for device-related information in EUDAMED under EU MDR (EU) 2017/745 and IVDR (EU) 2017/746. It groups medical devices sharing the same intended purpose, risk class, and essential design characteristics, and never appears in packaging barcodes.
How is the GS1 GMN check character pair calculated?
Per GS1 General Specifications Section 7.9.5, each character in the GS1 Company Prefix plus Model Reference (up to 23 characters from GS1 AI set 82) is mapped to its reference value (0 to 81), multiplied by prime weights starting at 2 from right to left, summed modulo 1021, and split into two check characters excluding 0, O, 1, and I.
How does the IFA GmbH Mod-97 Basic UDI-DI algorithm work?
IFA Basic UDI-DIs begin with the IFA Product Registration Agency prefix (11) followed by the manufacturer/model code and a 2-digit Modulo 97 check number (computed per ISO/IEC 7064 Mod 97,10 on numeric digits or alphanumeric mapping).
Can I validate a spreadsheet list of Basic UDI-DIs at once?
Yes. Paste multiple Basic UDI-DI strings (one per line) into the calculator above, click Calculate, and export the full verification table as a CSV file for your technical documentation.